AP22 Speedy IF
The AP22 Speedy IF processor is manufactured by DAS and CE marked with respect to the In Vitro Diagnostic Devices Directive (98/79/EC). The AP22 Speedy IF processor was launched in the UK in January 2006 as part of Cambridge Life Sciences’ innovative diagnostic solution.
The AP22 Speedy IF processor is as an open system capable of running a variety of microplates ELISA and immunofluorescence assays (IFA). The flexibility of the AP22 Speedy IF processor enables the user to run a wide range of different assays or combination of assays to accommodate the laboratory workload.
Up to eight different ELISA assays can be run at the same time, providing the assays use no more than two different wash buffers and or two different sample diluents.
For IFA, the AP22 Speedy IF processor can run up to sixteen slides of any different combination of substrates and layout. By utilising the predilution tray (92 +108 positions), samples can be run at different screening dilutions (automatically selectable from the software) and different titration protocols are easily accommodated.
Specifications |
Description |
Principle |
Automated processor to carry out:
ELISA tests on 96 wells micro-plates
IFA tests on glass slides |
Maintenance |
Daily maintenance if performed manually by the operator |
Operating system |
Windows 98 - XP |
Power requirements |
230 or 115 VAC, 50-60 Hz, 500W |
Space requirements |
79 x 61 x 65 cm |
Weight |
77Kg |
ELISA Technical Specification
Specifications |
Description |
Sample tubes |
80 x 12 mm. 80 x 2mL screw cap tubes.
Customised racks are available for other tube sizes |
Predilution tubes |
92 positions |
Reagent racks |
8 separate reagent racks hold up to 8 different sets of calibrators, controls and reagents |
Reservoirs |
1 x waste tank, 1 x distilled water, 2 x separate wash buffer bottles with level sensors |
Microtitre plates |
2 standard 12 x 8 well plates with programmable incubation from ambient to 450C. |
Plate washing |
Single manifold strip washer with 2 rows of 8 needles, one set for dispensing the other for aspirating. Automatic height adjustment for different types of microplate |
Reading system |
8 independent photometric channels with mono and dual wavelength |
Interferential filters |
2 supplied as standard (450 nm, 630 nm), option for up to 4 |
Reading capability |
From 0.000 to 3.000 OD, accuracy ±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD |
Dispensing |
8 μL to 1000 μL by steps of 1 μL, precision ±1% with 100 μL |
IFA Technical Specification
Specifications |
Description |
Sample tubes |
80 75mm x 12 mm or 80 2mL screw cap tubes.
Customised racks are available for other tube sizes |
Predilution tubes |
92 positions + 108 positions (optional) |
Reagent racks |
1x removable reagent rack, positions for 12 prediluted controls, 4 conjugates (or 3 conjugates and 1 Evans Blue), 1 sample diluent and 1 needle wash solution |
Reservoirs |
x waste tank, 1 x distilled water, 1 x PBS wash with level sensors |
Needle system |
2 needles on a single arm (one dispense, one aspirate) with independent movement |
Dispensing |
From 8 μL to 1000 μL by steps of 1 μL, precision ±1% with 100 μL |
Sample predilution |
20 seconds for a 1/100 (10 μL to 1000μL) |
Sample dispense |
9 seconds per well (up to 1000 μL). CV < 2% @ 100μL |
Reagent dispense |
38 seconds for a 10 well slide @ 20 μL per well (including rinse). CV < 2% @ 100 μL |
Slide rack |
1 x 16 slide removable rack |
Sample detection |
Can detect down to 150 μL sample volume |
Slide washing |
Flux system utilising simultaneous use of dispense and aspirate probes |
Additional options
Features |
Description |
Barcode reader |
hand held device |
Predilution tray |
108 positions |
|